Absorbance-to-Concentration (Beer-Lambert)
The current ledger marks this instrument verified clean within the source-bound verification corpus.
Inspectable records
Ledger state
The current ledger marks this instrument verified clean within the source-bound verification corpus.
- Area
- Biology, Medicine & Pharmacology (educational estimates only — every panel labeled not-medical-advice)
- Tranche
- tranche-5
- Evidence ID
- t5-157
- Ledger rows
- 1
No ledger note is attached.
Review findings
The formula A = ε·l·c and rearrangement c = A/(ε·l) are standard for molar absorptivity with compatible units. Verdict is NEEDS_REVIEW because the input specification does not explicitly define the units or required positivity constraints, and optional extinction-coefficient presets require wavelength and experimental-condition metadata.
Formula and scope findings
No additional record
Unit assumptions
- Absorbance A is dimensionless.
- The extinction coefficient ε is molar absorptivity in L·mol⁻¹·cm⁻¹ (equivalently M⁻¹·cm⁻¹).
- Path length l must be in centimeters when ε is expressed per centimeter.
- The output concentration c is in mol·L⁻¹ (M).
- The analyte, wavelength, solvent, pH, temperature, and other conditions must match the selected ε value.
Boundary hardening
BREAK_CLAIMED · SELF_FLAGGED
A zero extinction coefficient or zero path length produces an undefined division and must be rejected. The specification also provides no numeric validity threshold or alternate formula branch, and path-length units can cause a 100-fold concentration error if interpreted inconsistently with the extinction-coefficient units.
Errata and corrections
No applied erratum is attached to this unique-name record.
Inspectable example
4 review vectors are recorded. One source-bound sample is shown below.
{
"inputs": {
"absorbance": 0.5,
"molar_extinction_coefficient": 15000,
"extinction_coefficient_units": "M^-1 cm^-1",
"path_length": 1,
"path_length_units": "cm"
},
"expected_outputs": {
"concentration": 0.0000333333333333,
"concentration_units": "M"
},
"working": "c = A/(ε·l) = 0.500/(15,000 × 1) = 0.500/15,000 = 0.0000333333333333 M = 3.33333333333 × 10^-5 M."