Electrical Safety / Shock Current Reference mechanism
The inspectable method record for Electrical Safety / Shock Current Reference, kept separate from implementation and qualification.
Model scope
needs review because the claimed IEC 60479 comparison is underspecified and the supplied algorithm omits the standard's waveform, frequency, current path, and duration-dependent assessment. The four vectors are valid only for the explicitly stated Ohmic, 50/60-Hz RMS AC approximation and the simplified educational boundaries. Retain the educational-only warning and require lockout/tagout and applicable electrical codes rather than using this instrument for live-work or safety-certification decisions.
- I=V/Rbody is correct for an Ohmic body-resistance approximation, with voltage and resistance expressed in compatible units.
- The specification does not identify AC versus DC, frequency, waveform, current path, contact area, or whether the voltage is RMS or peak; IEC 60479-1 thresholds depend on these conditions.
- Fixed current bands are not a complete IEC 60479-1 method: the standard uses current-duration zones and physiological effects, including fibrillation-probability zones, rather than a universal hazard classification based only on current.
- The commonly used simplified 50/60-Hz AC boundaries of 0.5 mA, 10 mA, and 30 mA are educational shorthand and must not be labeled as a full IEC 60479-1 compliance calculation.
- Duration is an input but is ignored by the stated classification algorithm; this is technically inadequate for the claimed current-duration threshold reference.
- The claimed 'touch-voltage estimate' is actually just the supplied voltage unless a separate voltage-drop or contact-voltage model is specified.
Units and conventions
- Voltage is touch voltage in volts RMS for a sinusoidal 50/60-Hz AC source.
- Body resistance is in ohms and represents the complete hand-to-hand or hand-to-foot current path; it is treated as constant during the contact.
- Current is calculated as amperes by I=V/R and also reported in milliamperes using 1 A=1000 mA.
- The test classification uses the simplified educational convention: AC-1 for I≤0.5 mA, AC-2 for 0.5 mA<I≤10 mA, AC-3 for 10 mA<I≤30 mA, and AC-4 for I>30 mA.
- The simplified classification is evaluated at the stated current and does not claim to reproduce the full IEC 60479-1 current-duration zones.
- A zero-resistance input is invalid because it causes division by zero; zero voltage with positive resistance is valid and produces zero current.
Inspectable vector
One of 4 source-bound review vectors:
{
"inputs": {
"voltage_V": 120,
"body_resistance_ohm": 100000,
"duration_s": 1,
"waveform": "50/60-Hz sinusoidal AC assumed"
},
"expected_outputs": {
"touch_voltage_V": 120,
"estimated_current_A": 0.0012,
"estimated_current_mA": 1.2,
"hazard_band": "AC-2: above 0.5 mA through 10 mA in the simplified educational convention"
},
"reference_data": {
"perception_boundary_mA": 0.5,
"let_go_boundary_mA": 10,
"severe_contraction_boundary_mA": 30,
"source_convention": "Widely tabulated standard value, simplified 50/60-Hz AC educational shorthand; not a complete normative IEC 60479-1 table",
"classification_rule": "AC-1 if I≤0.5 mA; AC-2 if 0.5 mA<I≤10 mA; AC-3 if 10 mA<I≤30 mA; AC-4 if I>30 mA"
},
"working": "I=V/R=120 V/100000 Ω=0.0012 A. Convert to milliamperes: 0.0012 A×1000=1.2 mA. Since 0.5 mA<1.2 mA≤10 mA, the simplified classification is AC-2. The supplied touch voltage is 120 V."
}Failure boundary
Known limit · flagged by the author
A zero body-resistance input produces an undefined division result and must be rejected. The reference profile is also under-specified because the actual IEC 60479-1 table, waveform, frequency, current path, and duration-dependent thresholds are not supplied; the branch classifications above therefore use explicitly labeled illustrative thresholds rather than constituting a safety certification.