Absorbance-to-Concentration (Beer-Lambert) implementation
A source-bound implementation contract for Absorbance-to-Concentration (Beer-Lambert); deployment and executable qualification remain separate.
Binding contract
This record describes what an executable surface must bind and validate. It does not claim that a public calculator, package, or production route is deployed.
- Record owner
- Guth Instruments
- Lifecycle
- verified-clean
- Evidence ID
- t5-157
- Effect
- Computation or source lookup only
Input and validation requirements
A zero extinction coefficient or zero path length produces an undefined division and must be rejected. The specification also provides no numeric validity threshold or alternate formula branch, and path-length units can cause a 100-fold concentration error if interpreted inconsistently with the extinction-coefficient units.
Required corrections
No binding erratum is attached to this unique-name record. Implementers must still preserve the declared units, domain, and failure behavior.
Inspectable interface fixture
The sample below is evidence for interface and validation tests, not a production implementation.
{
"inputs": {
"absorbance": 0.5,
"molar_extinction_coefficient": 15000,
"extinction_coefficient_units": "M^-1 cm^-1",
"path_length": 1,
"path_length_units": "cm"
},
"expected_outputs": {
"concentration": 0.0000333333333333,
"concentration_units": "M"
},
"working": "c = A/(ε·l) = 0.500/(15,000 × 1) = 0.500/15,000 = 0.0000333333333333 M = 3.33333333333 × 10^-5 M."
}Other records
Sources and evidence
These pointers bind the page to evidence. They do not grant authority or collapse verification into deployment.
VERIFIED_LEDGER.jsonevidence-tranche5/tranche5-results.json