EMC/EMI Near-Field Distance & Wavelength Calculator implementation
A source-bound implementation contract for EMC/EMI Near-Field Distance & Wavelength Calculator; deployment and executable qualification remain separate.
Binding contract
This record describes what an executable surface must bind and validate. It does not claim that a public calculator, package, or production route is deployed.
- Record owner
- Guth Instruments
- Lifecycle
- verified-clean
- Evidence ID
- t5-061
- Effect
- Computation or source lookup only
Input and validation requirements
The zero-frequency vector exposes an undefined division-by-zero case that must be rejected. The unit probe also exposes a severe convention hazard: ignoring the D unit changes the Fraunhofer result by four orders of magnitude.
Required corrections
No binding erratum is attached to this unique-name record. Implementers must still preserve the declared units, domain, and failure behavior.
Inspectable interface fixture
The sample below is evidence for interface and validation tests, not a production implementation.
{
"inputs": {
"frequency_hz": 100000000,
"D_m": 0.3
},
"expected_outputs": {
"wavelength_m": 2.99792458,
"reactive_scale_m": 0.477134516,
"fraunhofer_boundary_m": 0.060041537
},
"working": "λ = c/f = 299792458 / 100000000 = 2.99792458 m. Reactive scale = λ/(2π) = 2.99792458 / 6.283185307 = 0.477134516 m. Fraunhofer boundary = 2D²/λ = 2(0.3²)/2.99792458 = 2(0.09)/2.99792458 = 0.18/2.99792458 = 0.060041537 m."
}Other records
Sources and evidence
These pointers bind the page to evidence. They do not grant authority or collapse verification into deployment.
VERIFIED_LEDGER.jsonevidence-tranche5/tranche5-results.json